Jared,

If you could merge the changes I made to the calibration data and then update the file on p8portal, that would be great.

Thanks,
Micah

@buuck

buuck commented Aug 17, 2012

Copy link
Copy Markdown
Author

Actually, hold off on that for now. I want to make sure that the error is in the calibration, and not some error in dripline reading the resistances incorrectly.

@buuck

buuck commented Aug 17, 2012

Copy link
Copy Markdown
Author

Ok, I checked out the resistances of the cernox sensors, and it looks like they all agree with what the muxer is reading, which would imply that the error is in the calibration. This pull request should fix it.