In 2024, the Centers for Disease Control and Prevention (CDC) estimated that over 105,000 people in the U.S. died from drug overdoses. Opioid overdoses occurs when there are so many drugs in the body that the individual becomes unresponsive and cannot breathe on their own. After 3-5 minutes without oxygen, brain damage begins, soon followed by death. Survival in the event of an opioid overdoses depends solely on getting oxygen to the brain. There is however a life saving drug, naloxone, that if administered within this critical three-minute period can reverse the effects of an overdose.

Detecting an overdose can be difficult, with symptoms including unresponsiveness and slowed, shallow, or stopped breathing. It is often challenging to distinguish if a person is sleeping or overdosing in public settings.

Numerous efforts have been made by the White House and public health organizations throughout the opioid epidemic. A barrier to many of the previous approaches is that many individuals who would benefit from them are not in the mental or physical state to utilize them.

Our product aims to address these shortcomings and introduce a new technology to automate overdose detection.

NaloxSAVER Product

NaloxSAVER is a web interface program that incorporated thermal infrared and RGB camera footage to track individuals breathing patterns and alert of possible overdoses in public spaces.

NaloxSAVER is designed to be installed in public spaces, such as train cars or public libraries. The system includes a tablet interface with a live video stream interface, a naloxone dispenser located next to the tablet, and two stacked cameras (thermal and RGB) to monitor passengers' breathing patterns. Each passenger's nose region and average temperature are tracked every second and stored internally. By detecting temperature changes between frames, the system can identify individuals experiencing a potential overdose and trigger an alarm.

The interface features a large screen with a live video feed, a "turn off alarm" button, and a "Narcan instructions" button. When an overdose is detected, the screen perimeter will blink red and white, accompanied by an alarm to alert bystanders. Users can click the "Narcan instructions" button to view an informational video from the CDC on administering the drug.

Software

For our thermal infrared camera, we used Hikmicro's Pocket2 which has a resolution of 49,152 pixels with a 50° × 37.2° field of view. For the RGB camera, we used a full HD 1080p compact webcam. We aligned our thermal and RGB cameras on the same plane and matched the zoom levels to our best ability. This alignment was achieved through adjustments in x-offset, y-offset, and zoom. Both cameras are kept in a static position to ensure these variables remain constant in each frame.

For facial recognition, we used Google's open-source model, Mediapipe. Mediapipe plots six key landmarks on the face (eyes, ears, nose, and mouth), which allowed us to extract only the nose point. We then overlaid these coordinates onto the thermal camera footage to calculate temperature data via our pixel color conversion equation.

To make NaloxSAVER technology more versatile and applicable to a public space, we created a "person" class within our program that allowed the system to track multiple people and store their temperature history simultaneously. Instances of the person class are maintained and updated relative to the subjects last detected position. As long as a person stays within 75 pixels of their last detected location, the same instance will be update. If the person was not detected for more than 10 seconds (ie. moved outside 75 pixels), their instance was deleted from the people list to increase the overall efficiency of the program.

To detect an overdose, we measure the average temperature around a person's nose once per second and store this data in memory. We compare each new temperature measurement with the previous one. If the temperature difference is less than 0.125°C, it indicates no breathing was detected. If this condition persists for seven consecutive seconds, we determine that the person is having an overdose and trigger the alarm. However, if the temperature difference exceeds 0.125°C at any point, the counter resets to 0, because breathing has been detected then the overdose detection process starts over.

Hardware

In addition to our software detection interface, we created 3D models to hold the camreras in a static position and a naloxone dispenser. Our naloxone dispenser was inspired by a candy dispenser and fit to the dimensions of Narcan nasal spray units The tablet holder dimensions are based on an Amazon Fire tablet, and the dispenser is designed to fit ten Narcan nasal spray units.

Testing

To test the accuracy of our product, we performed six rounds of breathing exercises at three different distances (1ft, 3ft, 5ft). The tests are structured as follows:

We recorded the results in a binary confusion matrix and calculated accuracy along with pixel coverage. The confusion matrix also allowed us to easily identify type I and II errors. Testing was consistently done on the same participant to minimize variability.

Results

1 Foot Away

Of 36 total trials, we recorded 18 true negatives, 16 true positives, 2 false positives, and 0 false negatives. This gave us a 94.4% overall accuracy rating. Our precision rate was 88.9%, resulting in a 10% type I error frequency. Our recall rate was 100% with 0% type II error. At a 1 foot distance, the individual's face covers approximately 66,706 pixels, which is around 20% of the total frame.

3 Feet Away

With the individual 2 feet from the camera, we recorded 15 true negatives, 12 true positives, 2 false positives, and 6 false negatives. This gave us a 77.1% overall accuracy rating. Our precision rate was 66.7%, resulting in a 28.6% type I error frequency. Our recall rate is 80% with a 20% type II error frequency. At 3 feet, the individual's face covers 11,641 pixels, which is about 3.7% of the total frame.

5 Feet Away

At 5 feet distance from the camera the system detected 16 true negatives, 10 true positives, 8 false positives and 2 false negatives. This produced a 72.2% accuracy rate. Our precision rate is 55.6%, resulting in an 33.3% Type I error frequency. Our recall rate is 83.3% with a 16.7% Type II error frequency. At 5 feet, the individuals face was covering 7,657 pixels which is about 2.5% of the total frame.

Disscusion

Our results demonstrate a marked decrease in accuracy as the distance between the subject and the camera increases. The accuracy decreases as distance increases, as seen from 94% at 1 foot to 75% at 3 feet and finally to 72% at 5 feet. This trend can be linked to the reduction in pixel coverage of the subject's face as the distance increases. The accuracy decrease from 3 feet to 5 feet is less dramatic, indicating a potential saturation point in accuracy deterioration. At this range, the system's accuracy reaches a lower bound, suggesting that further reductions in resolution have diminishing returns. The stabilization in results between 3 feet and 5 feet show that the system's accuracy has already been compromised by reduced detail from pixel coverage, and additional decreases will have progressively smaller impacts on accuracy performance.

To maintain a 94% accuracy rate at increasing distances we have create a recommendation for camera resolution. At 3 feet, we believe a minimum of a 541,249 pixel resolution thermal camera will maintain accuracy ratings. At 5 feet, we recommend upgrading to around 798,720 pixel thermal camera.

Limitations and Future Work

Our system provided a comprehensive proof of concept that also revealed several key limitations that should be addressed in future work.

Pixel Coverage and Saturation

One significant limitation is the impact of distance on pixel coverage and detection accuracy. Our findings indicate a notable decrease in accuracy as the subject moves further from the camera, primarily due to a reduction in the number of pixels covering the face. This was likely due to the resolution of our thermal camera and capability to perform at high levels with greater distances.

To address the camera limitations, upgrading the system’s resolution with higher-resolution cameras could reduce the impact of less pixel coverage at greater distances.

Enviroment

This study was conducted in a controlled environment, which does not reflect potential real-world implementations. Factors such as lighting variations, external impacts on temperature (ie. doors opening), multiple subjects and subject movement were controlled so our results are exclusively evaluating accuracy.

Evaluating the system’s usability and effectiveness in real-world settings will be crucial for its successful deployment and ensuring it meets the needs of end users. This may include testing different lighting conditions, using multiple subjects in a frame, and having subjects move around the frame. Understanding how these factors affect pixel coverage and accuracy could educate future adjustments to the system’s design.

System Training

Our product used frame by frame temperature detection using a pixel color to temperature equation. While we use MediaPipe which is pre-trained for facial detection, our breathing detection system is very adaptive. A limitation to this approach is that our system is not trained to understand breathing patterns and have the ability to mediate errors.

To address this, future work could explore training data sets to increase accuracy and reliability. This would allow for types of breathing patterns to be better handled and could decrease type II error frequency. It would also allow for better testing effectiveness and make strides towards deployment measures.

How to Run / Study Replication

This section will give instructions on how users can run and setup an environment to test our product and conduct further research.

Prerequisites

Before anything, you must make sure that you are running the correct versions of the necessary API's and libraries that we used. To do this, you should first make sure that your coding IDE is fully up to date and that your laptop is capable of running and performing computations on multiple cameras at the same time. This experiment was only able to be ran on Windows 11 devices (likely Windows 10 as well) and was ran on intel i7 10th and 9th gen mobile processors. You must also have Python 3.10.0 or newer to run this project. Future Python versions may not be supported, but this is unlikely for a long time.

Once you are sure that your machine is capable of running this experiment, make sure that you are in the same working directory as this project and make sure that the selected interpreter is the correct python version. Also, make sure that you have added python to your environment PATH variables. There are many tutorials online on how to do this, this is the one we used: https://realpython.com/add-python-to-path/. Once this step is completed, you can run the requirements.txt file in the terminal with this command: "$ pip install -r requirements.txt". There may be some complications, I suggest ChatGPT if there are any weird errors as they may be specific to your computer and file structure.

Camera Setup

Our product uses a Hikmicro Pocket2 thermal infrared camera and a 1080p RGB webcamera that both plug into our devices VIA USB. This connection allows us to stream live footage from both of these cameras simultaneously, and overlay them correctly. In order to replicate our study, you should make sure that the cameras are properly cropped and are displaying the exact same footage. You can do this by running our "camera_fit_test.py" file. You will most definitely need to make some adjustments because while the thermal camera will be easy to find online and to purchase, the RGB webcamera that we used is old and out of date.

In order to adjust your cameras (which should be in a static location so that neither of them move at all) upon running the camera fit test file and to have the exact same cropped live footage, first you need to make sure that the camera indices are correct. These will be either 0, 1, or 2 (maybe 3) and you can be assisted in finding these using "index_locator.py". Once you have adjusted the variables to display proper camera overlay, you can copy these variables into "app.py" and run the file.

Running our Experiment

Running our experiment is simple, once the app file is ran, you will see a locally hosted link in your terminal, usually pointing to an IP address. It could look as simple as this: "https://5000" but may vary from device to device. Once you follow this link in a browser of your choosing (we found firefox to work most smoothly), you will be greeted with our user interface and the live footage of yourself. From here, note that holding your breath for 7 seconds should detect an overdose. If an overdose is detected, you can press the reset button at the bottom of the screen to revert the program back to its original state. Our experiment tested the effectiveness of our program and setup, we hope to see more adjustments made by anyone willing to do further research!