ChemoPress prevents chemotherapy-induced peripheral neuropathy (CIPN) by applying controlled compression to the patient's hand during oxaliplatin infusion.
- Arduino Uno
- L298N Motor Drivers
- Adafruit MPRLS Pressure Sensor (I2C)
- 3x DC Air Pumps (6V-12V)
- Brass Air Release Valve
- System modeled as FOPTD: P(s) = Ke^(-Ls) / (τs+1)
- PI controller tuned using Cohen-Coon method
- Target pressure range: 20-33 hPa
- /code — Arduino sketches
- /hardware — Wiring and pin configuration
Kerdsaeng et al., ECTI-CON 2025, IEEE "Modeling of Pressure Control in Compressive Gloves for Neuropathy Prevention"