R-Net Programmer Protocol

Complete documentation of the R-Net R-Net Programmer protocol used for reading and writing wheelchair configuration data.

Overview

The R-Net Programmer is a commercial device made by PG Drives Technology for configuring R-Net wheelchair electronics. This document describes the CAN bus protocol used by the programmer to communicate with the Power Module (PM) and other R-Net devices.

Official Name: The programmer protocol is officially called POP (Parameter Object Protocol), running on the REBUS transport layer. These names were identified through reverse engineering of DongleInterface.dll v3.0.0.0 (PG Drives Technology), which implements the PC-side programmer interface.

Protocol Characteristics:

  • CAN 2.0B @ 125 Kbps
  • Standard frames (11-bit IDs) for basic POP operations
  • Extended frames (29-bit IDs) for segmented POP transfers
  • Request/Response pattern
  • No encryption or authentication

Frame IDs

Frame ID Direction Description
0x78F Programmer → PM Programmer request frames
0x793 PM → Programmer PM response frames
0x7A0 Programmer → Bus Programmer presence announcement
0x792 Device → Programmer Device responses (slot 2)

Note: Frame ID 0x793 is shared between R-Net Programmer responses and Slot 3 parameter responses. Context determines interpretation.

Command Bytes

The first byte of frame data indicates the command type:

Request Commands (0x78F)

Byte Command Description
0x42 PRESENCE_REQUEST Heartbeat/presence polling
0x83 SET_ADDRESS Set memory address pointer
0x03 START_READ Begin read operation
0x43 START_WRITE Begin write operation
0x20 OPEN_PAGE Open parameter page
0x40 REQUEST_DATA Request data at pointer

Response Commands (0x793)

Byte Command Description
0x2F PRESENCE_ACK Heartbeat acknowledgment
0x4F READ_RESPONSE Read data response
0x0F WRITE_RESPONSE Write acknowledgment
0x8F COMPLETE Operation complete
0x41 ACK General acknowledgment
0x21 DATA_RESPONSE Data with value
0xC1 ERROR Error response

Programmer Handshake

The handshake establishes programmer presence on the bus and must be maintained throughout the session.

Step 1: Announce Presence

Send an empty frame to announce the programmer:

TX: 7A0#

No response expected. This frame alerts devices that a programmer is connecting.

Step 2: Establish Heartbeat

Send heartbeat request and wait for acknowledgment:

TX: 78F#422000FF00000000
RX: 793#2F2000FF00000000

Frame Breakdown (Request):

78F#42 20 00 FF 00 00 00 00
     │  │  │  │
     │  │  │  └── Device ID (0xFF = broadcast)
     │  │  └───── Reserved (0x00)
     │  └──────── Register (0x20)
     └─────────── Command: Presence Request (0x42)

Frame Breakdown (Response):

793#2F 20 00 FF 00 00 00 00
     │  │  │  │
     │  │  │  └── Device ID echo
     │  │  └───── Reserved
     │  └──────── Register echo
     └─────────── Command: Presence ACK (0x2F)

Step 3: Maintain Connection

The heartbeat must be sent periodically to maintain the connection:

Interval: ~200ms
TX: 78F#422000FF00000000
RX: 793#2F2000FF00000000

If heartbeats stop, the PM will exit programmer mode after a timeout.

Complete Handshake Sequence

Time     Frame
──────────────────────────────────────────────────
  0ms    TX: 7A0#                         ; Announce presence
100ms    TX: 78F#422000FF00000000         ; Heartbeat 1
110ms    RX: 793#2F2000FF00000000         ; ACK
300ms    TX: 78F#422000FF00000000         ; Heartbeat 2
310ms    RX: 793#2F2000FF00000000         ; ACK
  ...    (continue every 200ms)

Read Protocol

Reading data from device memory.

Frame Format

Set Address Request:

78F#83 00 01 FF XX XX 00 00
     │  │  │  │  └──┴── Address (16-bit, little-endian)
     │  │  │  └──────── Device ID (0xFF = broadcast)
     │  │  └─────────── Sub-command (0x01)
     │  └────────────── Reserved (0x00)
     └───────────────── Command: Set Address (0x83)

Start Read Request:

78F#03 00 01 FF 17 00 00 00
     │  │  │  │  │
     │  │  │  │  └── Read size/mode (0x17)
     │  │  │  └───── Device ID
     │  │  └──────── Sub-command
     │  └─────────── Reserved
     └────────────── Command: Start Read (0x03)

Read Response:

793#4F 00 01 FF DD DD DD DD
     │  │  │  │  └──┴──┴──┴── Data (4 bytes)
     │  │  │  └────────────── Device ID
     │  │  └─────────────────  Sub-command
     │  └────────────────────  Reserved
     └───────────────────────  Command: Read Response (0x4F)

Operation Complete:

793#8F 00 01 FF 00 00 00 00
     │
     └── Command: Complete (0x8F)

Read Sequence

To read 4 bytes from address 0x0100:

1. TX: 78F#830001FF00010000    ; Set address to 0x0100 (little-endian: 00 01)
2. TX: 78F#030001FF17000000    ; Start read
3. RX: 793#4F0001FFDDDDDDDD    ; Response with data
4. RX: 793#8F0001FF00000000    ; Complete

Reading Multiple Bytes

For reads larger than 4 bytes, repeat the sequence with incremented addresses:

; Read 8 bytes from 0x0100

; Bytes 0-3
TX: 78F#830001FF00010000      ; Address 0x0100
TX: 78F#030001FF17000000      ; Start read
RX: 793#4F0001FF[byte0-3]     ; Data
RX: 793#8F0001FF00000000      ; Complete

; Bytes 4-7
TX: 78F#830001FF04010000      ; Address 0x0104
TX: 78F#030001FF17000000      ; Start read
RX: 793#4F0001FF[byte4-7]     ; Data
RX: 793#8F0001FF00000000      ; Complete

Write Protocol

Writing data to device memory.

Frame Format

Set Address Request: (same as read)

78F#83 00 01 FF XX XX 00 00

Write Command:

78F#43 00 01 FF DD DD DD DD
     │  │  │  │  └──┴──┴──┴── Data to write (4 bytes)
     │  │  │  └────────────── Device ID
     │  │  └─────────────────  Sub-command
     │  └────────────────────  Reserved
     └───────────────────────  Command: Start Write (0x43)

Write Response:

793#0F 00 01 FF SS 00 00 00
     │           │
     │           └── Status byte
     └────────────── Command: Write Response (0x0F)

Write Sequence

To write DEADBEEF to address 0x0200:

1. TX: 78F#830001FF00020000    ; Set address to 0x0200
2. TX: 78F#430001FFDEADBEEF    ; Write data
3. RX: 793#0F0001FF41000000    ; Write ACK (0x41 = success)
4. RX: 793#8F0001FF00000000    ; Complete

Error Handling

Error Response

When an operation fails, the PM responds with error code 0xC1:

793#C1 81 00 00 28 00 00 00
     │  │        │
     │  │        └── Error code
     │  └─────────── Register that failed
     └────────────── Command: Error (0xC1)

Common Error Codes

Code Meaning
0x28 Address not found
0x00 General error

Timeout Handling

  • Read timeout: 500ms
  • Write timeout: 500ms
  • Heartbeat timeout: 200ms
  • Session timeout: ~2000ms without heartbeat

Observed Traffic Patterns

From Captured Programmer Session

Traffic captured from an actual R-Net Programmer session shows this pattern:

; Initial handshake
7A0#                              ; Programmer announce
78F#422000FF00000000              ; Heartbeat
793#2F2000FF00000000              ; ACK

; Periodic status frames
0300:00400#1510                   ; Status: connected
0300:00400#1515                   ; Status: active

; Continuous heartbeat during operation
78F#422000FF00000000              ; Heartbeat (every 200ms)
793#2F2000FF00000000              ; ACK

Multi-Device Query

The programmer can query multiple devices:

78F#422000FF00000000              ; Broadcast query (FF)
792#2F2000FF00000000              ; Slot 2 response
793#2F2000FF00000000              ; Slot 3 response

Memory Map

Based on analysis of .R-net configuration files:

Address Range Content
0x0000-0x001F Header / File type
0x0020-0x015F Device configuration blocks
0x0160-0x020F Profile names (20 chars each)
0x0210-0x0FFF Profile configurations
0x1000+ Mode maps and parameter tables

Known Profile Offsets

0x0160: "Drive" (20 bytes)
0x0174: "Seating" (20 bytes)
0x0188: "L33tBT" (20 bytes)
0x019C: "Mouse Control 1" (20 bytes)
...

Implementation Notes

Timing Requirements

  • Minimum inter-frame gap: 1ms
  • Heartbeat interval: 150-250ms (200ms recommended)
  • Response timeout: 500ms
  • Maximum burst rate: ~10 frames/100ms

Device ID Filtering

  • 0xFF = Broadcast (all devices respond)
  • 0x00 = PM (Power Module)
  • 0x01 = JSM (Joystick Module)
  • 0x02-0x0E = Other devices
  • 0x0F = Reserved for programmer

Session Lifecycle

1. Open CAN socket
2. Send 7A0# (announce)
3. Start heartbeat thread (200ms interval)
4. Perform read/write operations
5. Stop heartbeat
6. Close socket

Security Considerations

The R-Net Programmer protocol has no security:

  1. No authentication - Any device can send programmer frames
  2. No encryption - All data transmitted in plaintext
  3. No integrity checking - No CRC beyond CAN layer
  4. Broadcast addressing - 0xFF queries all devices

This allows:

  • Configuration extraction (dump all memory)
  • Configuration modification (write arbitrary values)
  • Profile manipulation (change speed limits, etc.)
  • Potential safety bypass (if safety params are writable)

POP Transfer Codes

The POP (Parameter Object Protocol) defines the following transfer codes (TC) for data exchange between programmer and device:

Upload (Device to Programmer)

Code Name Value Description
TC_UL_REQ Upload Request 0x03 Programmer requests data from device
TC_UL_RES Upload Response 0x01 Device sends requested data
TC_ULS_REQ Upload Segmented Request 0x09 Segmented upload request (for large transfers)
TC_ULS_RES Upload Segmented Response 0x0A Segmented upload response

Download (Programmer to Device)

Code Name Value Description
TC_DL_REQ Download Request 0x05 Programmer sends data to device
TC_DL_RES Download Response 0x07 Device acknowledges download
TC_DLS_REQ Download Segmented Request 0x04 Segmented download request (for large transfers)
TC_DLS_RES Download Segmented Response 0x08 Segmented download response

Control

Code Name Value Description
TC_ABORT Transfer Abort - Abort current transfer
TC_END Transfer End - End of transfer sequence
TC_ILLEGAL Illegal Transfer - Invalid/unsupported transfer code

ODI Classes (Object Dictionary Interface)

The ODI (Object Dictionary Interface) classifies device memory regions into typed objects. The programmer uses ODI classes to address different memory types:

Class Code Name Description
0x0100 EEPROM Non-volatile configuration storage
0x0110 PORT I/O port registers
0x0120 RAM Volatile runtime data
0x0130 ROM Read-only firmware data
0x0140 ADC Analog-to-digital converter values
0x0200 SLOT Device slot configuration
0xFFFFFF SENTINEL End-of-list / uninitialized marker

ODI classes allow the programmer to access different memory types on the target device through a unified interface, rather than raw memory addresses.


POP Segmented Protocol (Extended CAN IDs)

For transfers larger than a single CAN frame, POP uses a segmented protocol with extended (29-bit) CAN IDs:

Extended Frame ID Construction

Request (Programmer to Device):

ID = 0x1E400000 | (node << 15) | (tc << 18)

Where:
  node = target device node number (0-F)
  tc   = transfer code (see POP Transfer Codes)

Response (Device to Programmer):

ID = 0x1E000000 | (node << 15) | (tc << 18)

Where:
  node = responding device node number
  tc   = response transfer code

Example

Segmented upload from node 1:

Request:  0x1E400000 | (1 << 15) | (0x09 << 18) = 0x1E648000
Response: 0x1E000000 | (1 << 15) | (0x0A << 18) = 0x1E288000

USB Dongle Interface

The R-Net Programmer connects to a PC via a USB dongle that bridges USB to CAN.

Hardware Details

Property Value
Part Number D50610
Interface Chip FTDI (USB-to-serial)
Device Description "RNet Dongle"

Framing Protocol

The USB dongle uses a byte-stuffed framing protocol:

Byte Name Value Description
STX Start of Text 0x02 Frame start delimiter
ETX End of Text 0x03 Frame end delimiter
DLE Data Link Escape 0x10 Escape character for byte-stuffing

Maximum frame size: 32 bytes

When data bytes match STX, ETX, or DLE, they are prefixed with DLE (byte-stuffing) to avoid framing ambiguity.

Raw Message Formats

V1 Format (21 bytes): Legacy message format for older programmer hardware.

V2 Format (24 bytes): Extended message format with additional fields for newer hardware.

Access Levels

The dongle's EEPROM stores access level credentials:

Level Privileges Description
OEM Full access Factory-level read/write to all memory regions
Dealer Restricted Limited to user-configurable parameters

Enforcement: Access levels are enforced by credentials stored in the dongle's EEPROM. OEM access grants strictly more privileges than Dealer access. The dongle presents its access level during the REBUS connection handshake.


Connection State Machine

The programmer connection follows a defined state progression:

CXTN_NONE ──> CXTN_INIT ──> CXTN_DONGLE ──> CXTN_CAN ──> CXTN_RNET ──> CXTN_UPLOAD
                                                                     └──> CXTN_DOWNLOAD
State Description
CXTN_NONE No connection, initial state
CXTN_INIT Connection initialization started
CXTN_DONGLE USB dongle detected and configured
CXTN_CAN CAN bus interface active (dongle talking to bus)
CXTN_RNET REBUS protocol established (heartbeat active, devices enumerated)
CXTN_UPLOAD Upload session active (reading data from device)
CXTN_DOWNLOAD Download session active (writing data to device)

The state machine progresses linearly from NONE through RNET. From RNET, the connection branches into either UPLOAD or DOWNLOAD mode depending on the operation. Failure at any stage returns to CXTN_NONE.


Timing Constants

Constants extracted from DongleInterface.dll:

Constant Value Description
QCK_TIMEOUT 1000ms Quick operation timeout
SEG_TIMEOUT 100ms Segmented transfer inter-frame timeout
DONGLE_TIMEOUT 500ms USB dongle response timeout
CANMSG_TIMEOUT 10000ms Maximum wait for any CAN message
USB_READ_TIMEOUT 250ms USB read operation timeout
READ_RETRY_COUNT 3 Number of read retries before failure
LATENCY_TIMER 1ms FTDI latency timer setting

Note: These values supplement the timing requirements in the existing Timing Requirements section. The SEG_TIMEOUT of 100ms is notably shorter than the standard heartbeat interval, reflecting the tighter timing needs of segmented bulk transfers.


Tool Usage

The rnet_programmer.py tool implements this protocol:

# Test handshake
./rnet_programmer.py handshake -v

# Enable programmer mode
./rnet_programmer.py enable

# Read 256 bytes from address 0x0000
./rnet_programmer.py read 0x0000 0x100

# Write data to address
./rnet_programmer.py write 0x0100 --data DEADBEEF

# Dump memory range
./rnet_programmer.py dump 0x0000 0x1000 -o config.bin

# Monitor programmer frames
./rnet_programmer.py monitor --filter 78F,793,7A0

References

  • DEFCON24 Wheelchair Hacking presentation (Stephen Chavez & Specter)
  • wireshark/programmer_dump_file_july2017.pcapng - Captured programmer read session
  • wireshark/programmer_write_file_july2017.pcapng - Captured programmer write session
  • wireshark/ics_write_config.pcapng - ICS configuration transfer
  • new/src/test.cpp - Captured programmer bus queries (lines 492-591)
  • ../reference/RNET_FRAME_DICTIONARY.md - Complete frame reference
  • DongleInterface.dll v3.0.0.0 (PG Drives Technology) - Source of POP/REBUS/ODI terminology, transfer codes, ODI classes, timing constants, USB dongle protocol, connection state machine, and access levels